Radiation dose to organs outside the radiotherapy treatment field can be significant and clinical interest. Therefore, in order to estimate peripheral doses (PD) during the stereotactic radiosurgery treatment with Leksell Gamma Knife for single isocenter treatment technique in particular as a part of the method can be described the all plan treatment in general. Knowledge of PD can be used effectively during the optimization of treatment planning in order to deliver a very accurate dose of radiation to well-defined target with minimal damage to surrounding healthy tissue and other organs. Moreover, PD can be limiting factor during the treatment of some non-malignant lesion or pregnant woman.
Alderson-Rando phantom was used for PD measurements. The
Dose of 40 Gy was delivered in a single fraction radiosurgery to midline hypothetical
target volume situated close to the center of calva. Treatment planning for
single target covered by 50% isodose of each collimator was performed with
Leksell Gamma Plan (ver. 5,32) treatment planning system. The treatment plans
for a single, 5, 10, 15, 20 and 25 shots, keeping the same position and the
same doses were produced. Automatic positioning system (APS) was used for
positioning of phantom during the treatment. The doses to different organs were
measured during treatment by thermoluminescence dosimeter (TL).
A simple exponential function can describe dependence of
measured PD on distance from target for all collimator size used. Close to the
target a scattered radiation within the patient was dominant, at distances
beyond 40 cm from the target this component becomes less predominated and less
dependence of the collimator size. While the PD distances beyond 40 cm was
mainly due to scattered radiation in the treatment room and leakage radiation.
Contributions to the PD due the transportation of patient during the
positioning were linearly increasing with the number of shots. DT does not
practically contribute to the prescribed dose in target and to the laterally
scattered radiation within patient. PD depends on collimator size and this
dependence is more significant close to the target where it is proportional
with the C3 and less significant far from the target where it is
proportional with C only but more dependence on the number of shots.
The number of shots as well as collimator size has influence
on the magnitude of PD. Reduction of number of shots can decrease the PD for
sites far from the target, on other hand the influence of collimator size is
more significant for organs close to the target. Weighting the effect of the
collimator size and number of shots can be used effectively during the
optimization procedure to choose the most suitable treatment plan.
Author (s) Details
Mustafa Majali
Federal Authority for Nuclear Regulation (FANR), PO Box 112021, Abu Dhabi,
UAE.
Please see the book here:- https://doi.org/10.9734/bpi/stda/v2/2420
No comments:
Post a Comment