Thursday, 10 April 2025

Plasmalyse – A Novel Plasma Sterilization System for Dental and Medical Applications | Chapter 26 | Plasmas Afterglows with N2 for Surface Treatments synthesis 2024

Reactors for medical instrument sterilization were studied by starting from 5-liters to 17-liters volumes for dental applications. The first challenge was to find low-cost elements to obtain a prize in the same order as a commercial autoclave. For that, a microwave source from a commercial microwave oven and the small vacuum pump of the autoclave were tested to build up a new plasma sterilizator, named Plasmalyse. Sterilization of 6 Log (10-6 spores survivors) was obtained after a treatment of 7 minutes at 100°C and 120 minutes at 30°C in comparison to 40 minutes at 105°C with the commercial autoclave. After that, Plasmalyse was transformed for the Hospital application, with a 135 liters reactor, connected to 4 microwave sources and with an efficient vacuum pump, to obtain a 6 Log spores inactivation at 60°C in a treatment time of 20 minutes. Moreover, the N-atom transmission and E-coli inactivation through hollow tubes were studied.

 

Author (s) Details

Sandrine Villeger
Department of Physics and Chemistry of Solids and Surfaces, Institut Jean Lamour, Nancy- Universite, CNRS, Parc de Saurupt, CS 14234, 54042 Nancy Cedex, France.

 

André Ricard
LAPLACE, Université de Toulouse, CNRS, INPT, UPS, 118 route de Narbonne, 31062 Toulouse Cedex 9, France.

 

Please see the book here:- https://doi.org/10.9734/bpi/mono/978-93-49473-93-5/CH3686.26